PICC Trainer — Troubleshooting

Weak flashback, leaks, catheter advancement problems, and when to replace a cartridge.

Updated 2026-09-22

Almost every problem with this trainer involves fluid level, air in the line, or a cartridge at end of life. Work through the checks in that order before concluding something is broken.

Check the reservoir first, every time

Low fluid level is a common fault. If the bag is below the refill line, order a replacement and retest before troubleshooting anything else.

Flashback problems

No flashback, or very weak flashback

  1. Check the fluid level. Low fluid is the usual answer.
  2. Confirm the clamp is open.
  3. Check the supply line for air and kinks. Re-prime per step 5 of the Setup Guide if you see bubbles.
  4. If level and line are both correct, the vessel cartridge has likely reached end of life. Replace it.

Bleeding is too vigorous

The bag is hanging too high or was overfilled. Lower it to 30 inches above the puncture site and confirm the fill level is at 500 mL, not above it.

Bleeding slows noticeably across a session

Expected behaviour as the bag drains. Refill at the refill line rather than waiting for the flow to become obviously weak - by the time it looks wrong, several students have already practiced on the wrong pressure.

Flow stops entirely

A closed clamp, a kinked line, or an airlock. Re-prime the line.

Leaks

Fluid appears away from the puncture site

Stop immediately and check the vessel cartridge luer lock. If it is flush and secure, the vessel has been punctured through its back wall. Replace the cartridge.

A leak inside the limb is not always visible

If fluid is pooling on the bench with no obvious source, close the clamp and inspect before continuing. Fluid tracking inside the limb saturates the material and shortens the life of everything it touches.

The vessel leaks after the catheter is withdrawn

The cartridge has reached end of life. The self-sealing layer stops closing before it stops accepting punctures, so this is the signal to replace it, not the point at which sticks become impossible.

The skin pad weeps around old puncture sites

Same cause, one layer up. Replace the skin pad.

Placement problems

The catheter will not advance

Usually the introducer is not fully inside the vessel lumen. Withdraw slightly, confirm flashback, and re-advance. If it happens again on a fresh attempt, check that the cartridge is not kinked at either end — a cartridge seated under tension narrows the lumen at the ends.

Depending on the size of PICC used, the catheter may get stuck or trapped in a connector or fornix created at the junction of the amber tubing and connector. In this situation, retract the catheter slightly, and try pushing again. If this issue frequently occurs, try using a smaller lumen PICC.

The vessel feels too soft, or offers no resistance

Either the reservoir is low, or the cartridge has been punctured enough times that the wall no longer holds pressure. Check the level first, then replace.

The skin pad tears or will not hold suture

End of life. Replace it. Suturing wears the pad faster than puncture does, so expect to replace pads more often if your sessions include securement practice.

When to replace versus when to report

Replace the consumable if the problem is a leak at a puncture site, a vessel that no longer reseals, or a torn skin pad. These are wear items and this is normal.

Contact us if the problem is anything structural — a cracked limb, a supply line fitting that leaks at the connection rather than at the vessel, or a new cartridge that fails on its first few sticks.

Email support@dvm-iq.com for any structural issues.

Next

For first-time setup and normal operation, see PICC Trainer — Setup.

Didn't solve it?

Email support@dvm-iq.com with your clinic name, the product, and what you've already tried.