PICC Trainer — Setup
Unboxing, filling the reservoir, and placing your first PICC.
This guide takes a new PICC Trainer from the carton to a first successful placement. Plan about ten minutes.
What's in the box
| Item | Quantity | Notes |
|---|---|---|
| Trainer limb | 1 | Serial number on the underside |
| Vessel cartridge | 1 installed | 3 spare |
| Skin pad | 1 installed | 1 spare |
| Reservoir bag, 500 mL | 1 | Contents may have settled during shipping, shake well |
| Supply line with clamp | 1 | Attach to bag prior to use |
Catheters, introducers, dilators, guidewires and suture are not included. Use your own clinical stock — expired inventory works and is the cheaper option.
Nothing here should be used on a live animal
The reservoir holds non-sterile simulated blood. It is not a drug, not a sterile product, and not a medical fluid of any kind. Keep the bag away from your clinical fluid shelf, and use a sharps container for introducer needles exactly as you would in a live procedure.
Setting up
- Place the limb on a flat, wipeable surface with a disposable, absorbable pad underneath. The trainer is designed to be used on a bench, not held in the hand.
- Ensure all luer lock connections are secure and free of leaks.
- Check the fluid level in the bag, if it falls below the refill line, pressures will not be accurate and a replacement should be ordered.
- Hang the bag so its outlet sits about 30 inches above the puncture site. This single measurement sets vessel pressure and matters more than it looks.
- Open the clamp and let fluid run until the supply line is clear of air, then close it. Air in the line is the most common cause of weak or intermittent flashback.
- Open the clamp again and press gently on the skin pad. Firm resistance of the vessel means the system is pressurised and you are ready.
Check the cartridge seal before you fill
The vessel cartridge (amber tubing) is seated during production but can loosen in transit. It should sit flush with no visible gap at either end. Reseat it before filling if it moves under light finger pressure - a loose cartridge leaks under pressure, and it will leak inside the limb where you cannot see it.
Your first PICC
- Place as directed by the manufacturer in a real patient.
Flashback is the check
If a student gets no flashback but the introducer is clearly in the vessel, the reservoir is low. Top it up rather than adjusting the hang height.
During a session
Refill the bag when it drops below the Refill Line. Pressure falls as the fluid level falls, and the range on the bag keeps flashback consistent.
Do not raise the bag to compensate for a low reservoir
Raising the hang height restores the bleed rate but overshoots physiologic pressure, and every subsequent stick reads wrong. Refill instead.
After a session
- Close the clamp and disconnect the fluid line.
- Flush the vessel cartridge with clean water and drain into the absorbable pad or sink.
- Wipe the skin pad and surrounding surface.
- Leave the limb uncovered until dry. Store at room temperature, out of direct sunlight.
Never store the trainer or the bag with fluid in it
Simulated blood left standing grows biofilm within a few days and permanently clouds the vessel. A cartridge stored wet is not covered by warranty.
Consumables and replacements
| Part | Typical life | Reorder |
|---|---|---|
| Vessel cartridge | 5-10 sticks | DVMiQ1002 |
| Skin pad | 10-20 sticks | DVMiQ1003 |
| Simulated blood concentrate (250mL) | 20-30 sticks | DVMiQ1004 |
| Simulated blood concentrate (500mL) | 50-75 sticks | DVMiQ1005 |
Email support@dvm-iq.com with your product number and quantity to reorder.
Next
If something isn't working as described here, see PICC Trainer — Troubleshooting.
Didn't solve it?
Email support@dvm-iq.com with your clinic name, the product, and what you've already tried.